Sliding wear characteristics of FDM-processed polylactic-acid in bovine blood serum

Authors

  • Budi Arifvianto Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia
  • A. T. Putra Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia
  • B. T. Prayoga Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia
  • R. Dharmastiti Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia
  • U. A. Salim Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia
  • M. Mahardika Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia
  • Suyitno . Centre for Innovation of Medical Equipments and Devices (CIMEDs) Department of Mechanical and Industrial Engineering, Faculty of Engineering, Universitas Gadjah Mada Jl. Grafika 2 Yogyakarta 55281, Indonesia

DOI:

https://doi.org/10.15282/jmes.13.4.2019.10.0466

Keywords:

wear, FDM, raster configuration, PLA, titanium, fused deposition modelling

Abstract

Fused deposition modelling (FDM) has so far been recognized for its reliability and simplicity for manufacturing of geometrically-complex polymeric materials with 3D printing technique. Recently, the studies concerning the properties of materials fabricated by using the FDM have been growing, including those related to their wear resistance which is considered of critical when a 3D-printed material must work sliding over the surface of another material during its application. Up to now, however, the influence of several FDM-printing parameters, including the raster orientations, on the wear resistance of the printed polymeric materials have not yet been fully understood. In this research, the influence of raster orientations on the wear resistance of FDM-processed polylactic-acid (PLA) materials in bovine blood serum were determined. A reciprocating pin-on-plate tribometer was used to evaluate wear resistance of a cylindrical PLA pin that slid over the surface of a commercially-pure titanium (Ti) plate. The results showed that the FDM-printed PLA pins with unidirectional raster orientations had higher wear factors (WF), i.e., ~1.5 ´ 10-4 mm N-1 m-1 than those with crossed raster orientations, i.e., ~0.8 ´ 10-4 mm N-1 m-1. Meanwhile, the WF values of Ti surface that slid against PLA pin with crossed raster orientations, i.e., ~0.09 ´ 10-4 mm N-1 m-1, were also significantly lower than those sliding with the pin having unidirectional raster orientation, i.e., 0.34 – 0.41 ´ 10-4 mm N-1 m-1. The result of analysis of the worn surface morphologies shows the indications of surface abrasion, fatigue and polymer film transfer as the possible wear mechanisms of both the FDM-processed PLA pin and its Ti countersurface material. On the basis of all the findings in this research, it can be concluded that the crossed raster orientation is preferable to be used as one of the parameters in printing of the wear resistant PLA rather than the unidirectional raster orientation.

References

Ventola CL. Medical applications for 3D printing: Current and projected uses. Pharmacy and Therapeutics. 2014; 39: 704-711.

Mertz L. Dream it, design it, print it in 3-D. IEEE Pulse, 2013; 4(6): 15-21.

Bogue R. 3D printing: The dawn of a new era in manufacturing? Assembly Automation. 2013; 33(4): 307-311.

Espalin D, Arcaute K, Rodriguez D, Medina F. Fused deposition modelling of patient-specific polymethylmethacrylate implants. Rapid Prototyping Journal. 2010; 16(3): 164-173.

Thomas DJ. 3D printing durable patient specific knee implants. Journal of Orthopaedics. 2017;14(1):182-183.

Li Z, Liu H. What we have achieved in the design of 3D printed metal implants for application in orthopaedics? Personal experience and review. Rapid Prototyping Journal. 2018; 24(8): 1365-1379.

Mohamed, OA, Masood, SH, Bhowmik, JL. Optimization of fused deposition modeling process parameters: A review of current research and future prospects, Advanced Manufacturing. 2015; 3:42–53.

Sood AK, Equbal A, Toppo V, Ohdar RK, Mahapatra SS. An investigation on sliding wear of FDM built parts. CIRP Journal of Manufacturing Science and Technology. 2012; 5(1): 48-54.

Equbal A, Sood AK, Toppo V, Ohdar RK, Mahapatra SS. Prediction and analysis of sliding wear performance of fused deposition modelling-processed ABS plastic parts. Proceedings of the Institution of Mechanical Engineers, Part J: Journal of Engineering Tribology. 2010; 224: 1261-1271.

Mohamed OH, Masood SH, Bhowmik JL. Analysis of wear behavior of additively manufactured PC-ABS parts. Materials Letters. 2018; 230: 261-265.

Borges RA, Choudury D, Zou M. 3D printed PCU/UHMWPE polymeric blend for artificial knee meniscus. Tribology International. 2018; 122: 1-7.

Zdero R, Guenther LE, Gascoyne TC. Pin-on-disc wear testing of biomaterials used for total joint replacements. In: Zdero R, editor. Experimental Methods in Orthopaedic Biomechanics, London: Academic Press, 2017, p 299-311.

Brown SS, Clarke IC. A review of lubrication conditions for wear simulation in artificial hip replacement. Tribology Transactions. 2006; 49: 72-78.

Prayoga BT, Suyitno S, Dharmastiti R. The wear behavior of UHMWPE against surface modified cp-titanium by thermal oxidation. Tribology in Industry. 2016; 38(4): 543-551.

Rajpurohit SR, Dave HK. Effect of process parameters on tensile strength of FDM printed PLA part. Rapid Prototyping Journal. 2018; 24(8): 1317-1324.

Attar H, Prashanth KG, Chaubey AK, Calin M, Zhang LC, Scudino S, Eckert J. Comparison of wear properties of commercially pure titanium prepared by selective laser melting and casting processes. Materials Letters. 2015; 142: 38-41.

Guezmil M, Bensalah W, Mezlini S. Tribological behavior of UHMWPE against TiAl6V4 and CoCr28Mo alloys under dry and lubricated conditions. Journal the Mechanical Behavior of Biomedical Materials. 2016; 63: 375-385.

Cooper JR, Dowson D, Fisher J. The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethylene. Clinical Materials. 1993; 14: 295-302.

Bahadur S. The development of transfer layers and their role in polymer tribology. Wear. 2000; 245(1-2) 92-99.

Downloads

Published

2019-12-30 — Updated on 2023-06-15

Versions

How to Cite

[1]
B. Arifvianto, “Sliding wear characteristics of FDM-processed polylactic-acid in bovine blood serum”, J. Mech. Eng. Sci., vol. 13, no. 4, pp. 5848–5861, Jun. 2023.

Similar Articles

<< < 3 4 5 6 7 8 9 10 11 12 > >> 

You may also start an advanced similarity search for this article.