Fabrication of aneurysm blood vessel silicone phantom using 3D printed semi-reusable mould
DOI:
https://doi.org/10.15282/jmes.19.4.2025.5.0853Keywords:
Blood vessel phantom, Silicon, 3D printing, Moulding, AneurysmAbstract
Aneurysms are vascular diseases with a low survival rate. Studying blood flow characteristics is essential to understand the disease, which would require experimentation on an anatomical phantom structure for simulating and measuring blood flow under controlled conditions. Conventional phantom fabrication often relies on single-use moulds, which are economically ineffective and time-consuming especially when replication is needed. In this study, we developed a method for fabricating an aneurysmal vascular phantom using a 3D-printed semi-reusable mould. The novelty lies in the reusable outer mould cope that was 3D-printed using polylactic acid (PLA), while the inner core was 3D-printed using water-soluble polyvinyl alcohol (PVA). A silicone-based elastomer, Sylgard 184 was used to create the vascular phantom because it approximates the mechanical properties of actual blood vessels. The silicone was injected onto the mould and the PVA core was dissolved in hot water to create a hollow structure. The resulting phantom is elastic and translucent, suitable for macro-scale flow visualization. It is also fabricated at a lower cost than those reported in previous studies. The semi-reusable mould concept enhances scalability by being low-cost and enables repeatable phantom production, offering an efficient way for research, training, and medical device testing.
References
[1] C. F. Wenceslau, CG McCarthy, S. Earley, S. K. England, J. A. Filosa, et al., “Guidelines for the measurement of vascular function and structure in isolated arteries and veins,” American Journal of Physiology–Heart and Circulatory Physiology, vol. 321, no. 1, pp. H77–H111, 2021.
[2] G. Toth and R. Cerejo, “Intracranial aneurysms: Review of current science and management,” Vascular Medicine, vol. 23, no. 3, pp. 276–288, 2018.
[3] J. Coles-Black, D. Bolton, and J. Chuen, “Accessing 3D-printed vascular phantoms for procedural simulation,” Frontiers in Surgery, vol. 7, pp. 1–12, 2021.
[4] E. Doutel, J. Carneiro, M. S. N. Oliveira, J. B. L. M. Campos, and J. M. Miranda, “Fabrication of 3D milli-scale channels for hemodynamic studies,” Journal of Mechanics in Medicine and Biology, vol. 15, no. 1, p. 1550004, 2014.
[5] J. Brunette, R. Mongrain, and J. C. Tardif, “A realistic coronary artery phantom for particle image velocimetry,” Journal of Visualization, vol. 7, no. 3, pp. 241–248, 2004.
[6] R. Jewkes, H. E. Burton, and D. M. Espino, “Towards additive manufacture of functional, spline-based morphometric models of healthy and diseased coronary arteries: In vitro proof-of-concept using a porcine template,” Journal of Functional Biomaterials, vol. 9, no. 1, pp. 1–17, 2018.
[7] P. F. Costa, H. J. Albers, J. E. A. Linssen, H. H. T. Middelkamp, L. Van Der Hout, R. Passier, et al., “Mimicking arterial thrombosis in a 3D-printed microfluidic in vitro vascular model based on computed tomography angiography data,” Lab on a Chip, vol. 17, no. 16, pp. 2785–2792, 2017.
[8] M. S. S. Hashuro, S. Tupin, N. K. Putra, K. Daibo, K. Inoue, T. Ishii, et al., “Development of an ultrasound phantom made of transparent material: Feasibility of optical particle image velocimetry,” Ultrasound in Medicine & Biology, vol. 49, no. 6, pp. 1385–1394, 2023.
[9] Y. Shimizu, N. K. Putra, and M. Ohta, “Reproduction method for dried biomodels composed of poly (vinyl alcohol) hydrogels,” Scientific Reports, vol. 8, no. 1, pp. 1–12, 2018.
[10] N. Kaneko, T. Mashiko, T. Ohnishi, M. Ohta, K. Namba, E. Watanabe, et al., “Manufacture of patient-specific vascular replicas for endovascular simulation using a fast, low-cost method,” Scientific Reports, vol. 6, no. 1, pp. 1–12, 2016.
[11] M. A. Cortez, R. Quintana, and R. B. Wicker, “Multi-step dip-spin coating manufacturing system for silicone cardiovascular membrane fabrication with prescribed compliance,” The International Journal of Advanced Manufacturing Technology, vol. 34, no. 7, pp. 667–679, 2007.
[12] Y. Shimizu, S. Tupin, C. Kiyomitsu, K. Kitamura, K. Takashima, and M. Ohta, “Development of a stereo dip-coating system for fabrication of tube-shaped blood vessel models,” Scientific Reports, vol. 10, no. 1, pp. 1–12, 2020.
[13] P. Cao, Y. Duhamel, G. Olympe, B. Ramond, and F. Langevin, “A new production method of elastic silicone carotid phantom based on MRI acquisition using rapid prototyping technique,” in Proc. 35th Annu. Int. Conf. IEEE Engineering in Medicine and Biology Society, Osaka, Japan, 2013, pp. 1–12.
[14] K. Jędrzejczak, A. Antonowicz, B. Butruk-Raszeja, W. Orciuch, K. Wojtas, P. Piasecki, et al., “Three-dimensionally printed elastic cardiovascular phantoms for carotid angioplasty training and personalized healthcare,” Journal of Clinical Medicine, vol. 13, no. 17, pp. 1–12, 2024.
[15] B. Guest, L. Arroyo, and J. Runciman, “A structural approach to 3D-printing arterial phantoms with physiologically comparable mechanical characteristics: Preliminary observations,” Proceedings of the Institution of Mechanical Engineers, Part H, vol. 236, no. 9, pp. 1388–1402, 2022.
[16] S. G. Yazdi, L. Huetter, P. D. Docherty, P. N. Williamson, D. Clucas, M. Jermy, et al., “A novel fabrication method for compliant silicone phantoms of arterial geometry for use in particle image velocimetry of haemodynamics,” Applied Sciences, vol. 9, no. 18, pp. 1–12, 2019.
[17] M. F. Hubakri, M. A. S. Zamri, M. N. A. Hamzah, R. A. Roslan, W. N. Wan Ab Naim, M. J. M. Mokhtarudin, “Design and development of a flexible test rig for biomedical engineering PIV experiments,” in Enabling Industry 4.0 through Advances in Manufacturing and Materials, Singapore, Springer, 2022, pp. 99–107.
[18] N. S. Che Amdilah, Y. S. Lau, Y. M. Liew, W. N. W. A. Naim, M. J. M. Mokhtarudin, “Computational fluid dynamics of patient-specific coronary artery during stent treatment,” in Engineering Technology International Conference (ETIC 2022), Kuantan, Malaysia, 2022, pp. 216–220.
[19] W. T. Tan, Y. M. Liew, M. J. Mohamed Mokhtarudin, S. Pirola, W. N. Wan Ab Naim, S. Amry Hashim, et al., “Effect of vessel tortuosity on stress concentration at the distal stent–vessel interface: Possible link with new entry formation through biomechanical simulation,” Journal of Biomechanical Engineering, vol. 143, no. 8, pp. 1–9, 2021.
[20] A. F. A. Ahmad Azahari, W. N. Wan Ab Naim, N. A. Md Sari, E. Lim, M. J. Mohamed Mokhtarudin, “Advancement in computational simulation and validation of congenital heart disease: A review,” Computer Methods in Biomechanics and Biomedical Engineering, vol. 1, no. 1, pp. 1–14, 2024.
[21] J. A. Hisam, M. Y. Salehudin, M. I. M. Lizah, “Fabrication of aneurysm biomodel using 3D printing technology,” in Advances in Intelligent Manufacturing and Mechatronics, Singapore, Springer, 2023, pp. 163–173.
[22] S. Tupin, K. M. Saqr, and M. Ohta, “Effects of wall compliance on multiharmonic pulsatile flow in idealized cerebral aneurysm models: Comparative PIV experiments,” Experiments in Fluids, vol. 61, no. 7, p. 164, 2020.
[23] S. F. Iftekar, A. Aabid, A. Amir, and M. Baig, “Advancements and limitations in 3D printing materials and technologies: A critical review,” Polymers, vol. 15, no. 11, p. 2519, 2023.
[24] R. Jafary, S. Armstrong, T. Byrne, A. Stephens, V. Pellegrino, S. D. Gregory, “Fabrication and characterization of tissue-mimicking phantoms for ultrasound-guided cannulation training,” ASAIO Journal, vol. 68, no. 7, pp. 940–948, 2022.
[25] J. Illi, A. W. Stark, C. Amstutz, J. Burger, P. Zysset, A. Haeberlin, et al., “Mechanical testing and comparison of porcine tissue, silicones, and 3D-printed materials for cardiovascular phantoms,” Frontiers in Bioengineering and Biotechnology, vol. 11, pp. 1–12, 2023.
[26] A. P. Ebrahimi, “Mechanical properties of normal and diseased cerebrovascular systems,” Journal of Vascular and Interventional Neurology, vol. 2, no. 2, pp. 155–162, 2009.
[27] Y. Wang, B. L. Tai, H. Yu, and A. J. Shih, “Silicone-based tissue-mimicking phantom for needle insertion simulation,” Journal of Medical Devices, vol. 8, no. 2, pp. 1–12, 2014.
[28] I. D. Johnston, D. K. McCluskey, C. K. L. Tan, and M. C. Tracey, “Mechanical characterization of bulk Sylgard 184 for microfluidics and microengineering,” Journal of Micromechanics and Microengineering, vol. 24, no. 3, p. 035017, 2014.
[29] M. E. Laughlin, S. E. Stephens, J. A. Hestekin, and M. O. Jensen, “Development of custom wall-less cardiovascular flow phantoms with tissue-mimicking gel,” Cardiovascular Engineering and Technology, vol. 13, no. 1, pp. 1–13, 2022.
[30] D. P. G. Nilsson, M. Holmgren, P. Holmlund, A Wåhlin, A. Eklund, T. Dahlberg, et al., “Patient-specific brain arteries molded as a flexible phantom model using 3D-printed water-soluble resin,” Scientific Reports, vol. 12, no. 1, p. 10172, 2022.
[31] M. Kaschwich, A. Dell, F. Matysiak, J. Bouchagiar, A. Bayer, M. Scharfschwerdt, F. Ernst, et al., “Development of an ultrasound-capable phantom with patient-specific 3D-printed vascular anatomy to simulate peripheral endovascular interventions,” Annals of Anatomy, vol. 232, p. 151563, 2020.
[32] H. Fu, H. Xu, Y. Liu, Z. Yang, S. Kormakov, D. Wu, et al., “Overview of injection molding technology for processing polymers and their composites,” ES Materials & Manufacturing, vol. 8, pp. 3–23, 2020.
[33] V. Carvalho, N. Rodrigues, R. Ribeiro, P. F. Costa, R. A. Lima, S. F. C. F. Teixeira, “3D-printed biomodels for flow visualization in stenotic vessels: An experimental and numerical study,” Micromachines, vol. 11, no. 6, pp. 1–12, 2020.
[34] S. C. H. Thian, Y. Tang, W. K. Tan, J. Y. H. Fuh, Y. S. Wong, H. T. Loh, et al., “The manufacture of micromoulds and microparts by vacuum casting,” The International Journal of Advanced Manufacturing Technology, vol. 38, no. 9, pp. 944–948, 2008.
[35] S. G. Yazdi, P. H. Geoghegan, P. D. Docherty, M. Jermy, and A. Khanafer, “A review of arterial phantom fabrication methods for flow measurement using PIV techniques,” Annals of Biomedical Engineering, vol. 46, no. 11, pp. 1697–1721, 2018.
Downloads
Published
Issue
Section
License
Copyright (c) 2025 The Author(s)

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.